NIDX-16
Non-Isolated Differential Voltage Module
16-Channel Non-Isolated Differential Voltage Module
The NIDX-16 input module provides 16 non-isolated differential voltage inputs and two analog (voltage) outputs for the Daxus® DXS-100 and SmartCorder® DDX-100 data acquisition systems. Connections for this module are made via screw terminals.
Non-Isolated Differential Voltage Input Module Key Benefits
- 16 voltage inputs (30 Vrms or ± 40 V DC max) and two voltage outputs (± 10V)
- Simultaneous sampling up to 20 kS/s/ch
- 16-bit resolution
- 0.04% accuracy (10V, 20V, and 40V attenuators)
- Built-in counter and timer functions
- Screw terminal or BNC connections with adaptor
Compatibility
Daxus® DXS-100 and SmartCorder® DDX-100 data acquisition systems
Typical applications
- Process monitoring and control
- Troubleshooting and repair
- Product verification and validation
+-Specifications
ANALOG INPUTS
- Channels Per Module: 16
- Connector: 8 wire screw terminals (8 connectors with 2 channels per connector)
- Input: Differential, DC coupled
- Bandwidth: 4.5 kHz (-3dB)
- Sample Rate: 20 kS/s
- A/D: 16 bit SAR (one per channel)
- Anti-Aliasing Filter: 4 pole Bessel
- Cold Start Drift: < 0.4% att (60 min.) – analog
- Off Ground Measurements: Yes
- Zero Suppression: Digital.
- Attenuator Ranges: 40, 20, 10, 5, 1, 0.5 Volt
- Minimum Span: 0.1 VFS
- Measurement Ranges: ± 40 V (40 VFS or 80 VFS w/ zero offset), ± 20 V (20 VFS or 40 VFS w/ zero offset), ± 10 V (10 VFS or 20 VFS w/ zero offset), ± 5 V (5 VFS or 10 VFS w/ zero offset), ± 1 V (1 VFS or 2 VFS w/ zero offset), ± 0.5 V (0.5 VFS or 2 VFS w/ zero offset)
- Max Rated Input: 30 Vrms or 40 VDC
- Max Common Mode Voltage: ± 60 V
- DC Accuracy (25°C): < 0.12% of attenuator (0.5V attenuator), < 0.10% of attenuator (1V attenuator), < 0.08% of attenuator (5V attenuator), < 0.04% of attenuator (10V, 20V and 40V attenuators)
- Overshoot: < 0.5%
- Intrinsic Noise (pk-pk): < 0.05% of attenuator + 1mV (0.5V, 1V and 5V attenuators), < 0.04% of attenuator (10V attenuator), < 0.03% of attenuator (20V and 40V attenuators)
- CMR at 60 Hz: Better than -70 dB (0.5 & 1V attenuators)
Better than -65 dB (5V, 10V, 20V and 40V attenuators)
- Min Input Impedance: > 500 Kohm
COUNTER TIMER
- Frequency Counter Capability: All channels. Software selectable.
- Counter Modes: Gated time frequency counter, cycle based frequency counter, pulse event, counter, gated pulse event counter, quadrature counter, pulse width detector, period width detector, duty cycle detector, edge separation detector.
- Frequency Ctr Range (menu): Up to 8 kHz
- Frequency Ctr Range (spec’d): 2 – 3 kHz
- Frequency Ctr Accuracy: ± 0.07% of Measurement + .002 Hz
- Min Counter Input Amplitude: 25% of span for freq and pulse counters, 90% of span for all other modes
- Pulse Counter Range: 4000000000 maximum. (16 bit display resolution)
- Pulse Width Accuracy: .002% of measurement + .00167% of span + 0.7 μs
- Pulse Width Range: 100 μs – 335000 μs
- Edge Separation Accuracy: .002% of measurement + .00167% of span + 0.7 μs
- Edge Separation Range: 100 μs – 5000000 μs
- Period Width Accuracy: .001% of measurement + .00167% of span + 0.7 μs
- Period Width Range: 100 μs – 100000 μs (10 Hz – 3 KHz)
- Duty Cycle Accuracy: .5% (Inputs in the 3 Hz – 3 KHz range with 5% – 95% duty cycles)
- Counter Timebase: 50 MHz
ANALOG OUTPUT
- Output Voltage Capability: Yes, two channels. One associated with ch 1-8 and the other ch 9-16.
- Output Voltage Range: Up to ± 10V
- Output Voltage Accuracy: ± 0.04V
- Output DAC Resolution/Speed: 12 bit / 1 MHz maximum
- Output Voltage Current: Up to 120 mA (Per output channel)
- Output Voltage Modes: DC, Arbitrary, Sine, Square and Pulse Train
- Arbitrary Memory Length: 32K per channel
Optional accessories
| SC-DAX Soft Carry Case for Daxus DXS-100, P/N: 42737000 |
| HC-DXS Hard Carry Case for Daxus DXS-100, P/N: 41047300 |
| SC-DDX Soft Carry Case for SmartCorder DDX-100, P/N: 41047200 |
Available services
- 1-Year extended warranty
- Factory calibration service