UNIV-6
Universal Module
Universal Input Module
The UNIV-6 is a 6-channel input module for voltage and strain/bridge measurements. Modules accept voltages up to 250 Vrms (banana jack inputs) and support full-bridge, half-bridge, and quarter bridge sensors. It provides maximum flexibility for acquiring different types of inputs with a single module.
Universal Module Key Benefits
- 6 voltage or strain/bridge inputs
- Simultaneous sampling at up to 800kS/s/ch
- 16-bit resolution
- 5 measurement ranges to maximize resolution
- Built-in counter and timer functions
- 250 VRMS or DC Cat II isolation
Compatibility
TMX®, TMX-18, and TMX-R data acquisition systems
Typical applications
- Troubleshooting
- Product testing
- Low channel count and mixed-signal applications
+-Specifications
Common Specifications
- Channels Per Module: 6
- Sample Rate: 800 kS/s/ch (400 KS/s/ch when using TMX-E or TMX-R)
- Rated Isolation: 250 VRMS or DC, Cat II (iso-common to chassis and other iso-commons)
- A/D: 16-bit SAR (one per channel)
- Frequency Counter Capability: Yes, first channel. Software selectable.
- Anti-Aliasing Filter: 4 pole Bessel
- Counter Modes: Gated time frequency counter, cycle based frequency counter, pulse counter, pulse width detector, period width detector, duty cycle detector
- Frequency ctr range (menu): Up to 120 kHz
- Frequency ctr range (spec’d): 2 – 100 kHz (48 Hz – 100 kHz for cycle based frequency counter)
- Frequency ctr accuracy: +/- 0.07% of Measurement +.002 Hz
- Min counter input amplitude: 25% of span for freq and pulse counters, 90% of span for all other modes
- Pulse counter range: 64000000 maximum. (16 bit display resolution)
- Pulse width accuracy: 0.7 μs +.00167% of span
- Pulse width range: 10 μs – 40000 μs
- Edge separation accuracy: .002% of measurement + .00167% of span + 0.7 μs
- Edge separation range: 25 μs – 5000000 μs
- Period width accuracy: .02% of measurement + .00167% of span + 1.0 μs
- Period width range: 5 μs – 90000 μs (11 Hz – 200 kHz)
- Duty cycle accuracy: .5% (Inputs in the 15 Hz – 10 kHz range with 20% – 80% duty cycles)
- Counter Timebase: 50 MHz
- Cold Start Drift: < 0.1% of attenuator (60 min.)
Single Ended Input
- Connector: Guarded banana jacks (red/black)
- Input: Single-ended, AC/DC coupled
- Anti-Aliasing Filter: 4 pole Bessel
- Bandwidth: 100 kHz (-3dB) (400V, 200V and 50V attenuators) 90 kHz (-3dB) (10V and 1V attenuators)
- AC Coupled 3dB Point: < 0.54 Hz
- Off Ground Measurements: Yes
- Zero Suppression: Yes
- Attenuator Ranges: 1, 10, 50, 200 and 400 Volt
- Measurement Ranges: +/- 400 V (400 VFS or 800 VFS w/ zero offset), +/- 200 V (200 VFS or 400 VFS w/ zero offset), +/- 50 V (50 VFS or 100 VFS w/ zero offset), +/- 10 V (10 VFS or 20 VFS w/ zero offset), +/- 1 V (1 VFS or 2 VFS w/ zero offset. 0.1V min span)
- Max Rated Input: 250 Vrms or DC, Cat II
- Max Transient Input: +/- 800 V peak (not to exceed 250Vrms)
- DC Accuracy (25°C): +/- 0.07% of attenuator
- Overshoot: < 1.0%
- Intrinsic Noise (pk-pk): < 0.08% of attenuator + .08% of span (400V through 10V atts) < 0.17% of attenuator + .07% of span (1V att
- IMR at 60 Hz: Better than -75 dB
- Min Input Impedance: > 1 Megohm
Differential Input
- Connector: 4 wire screw terminal
- Input: Differential, DC coupled
- A/D: 16 bit SAR (one per channel)
- Bandwidth: 50 kHz
- Measurement Ranges: +/- 1000 mV, +/- 500 mV, +/- 50 mV
- Max Transient Input: +/- 20V (no damage)
- Common Mode Voltage: +/- 3V
- Zero Suppression: Yes
- DC Accuracy (25°C): +/- 0.07% of attenuator
- Overshoot: < 1.0%
- Intrinsic Noise (pk-pk): < 0.08% of attenuator + .09% of span (1000 mV &500mV Atts) < 0.14% of attenuator + .08% of span (50 mV Att)
- Input Impedance: > 300 KΩ (150 KΩ balanced to isolated common)
- CMR at 60 Hz: > 85 dB
- Excitation: DC Voltage – adjustable, 0.1 to 10 V or 0.1 to 6 V for DC bridge measurements – 30 mA maximum
- Excitation Accuracy: 0.05 V voltage mode
† CMV specification applies only when pins other than the + and – inputs are being used. When the Differential / DC Bridge inputs are used simply as a differential input, the isolation mode voltage of 250 V applies
Optional accessories
| GL-40 General Use Lead Set, P/N: 13442000 |
| LC-40 Test Leads/Clips (Pair of test leads and pincer clips – one red and one black), P/N: 13441003 |
| LC-40S Test Leads/Spade Connectors (Pair of test leads with spade connector for #8 screw), P/N: 13441201 |
| ADP-4810 High Voltage Probe for up to 1000 Vrms, P/N: 25765000 |
| MAT-U* Universal Input Mating Connector (extra) 18 included at no charge, P/N: 25460104 |
| BNC-BAN-I Connector insulated Female BNC to standard insulated double Banana plug, P/N: 10532211 |
| HC-TMX Hard Transport Case for TMX , P/N: 41047010 |
| HC-TMXE Hard Transport Case for TMX and Expansion Chassis, P/N: 41047020 |
| SC-TMX Soft Carry Case for TMX, P/N: 41047000 |
Available services
- 1-Year extended warranty
- Factory calibration service