UNIV-4
Universal Input Module
Universal Module
The UNIV-4 is a 4-channel universal input module that measures voltage, temperature, strain, pressure, acceleration, current, and more. This module accepts voltages up to 250 Vrms and a wide variety of sensors including thermocouples, RTDs, bridge-based sensors, and accelerometers. Ideal for mixed-signal applications, the UNIV-4 provides maximum flexibility for acquiring different types of inputs with a single module.
Key Benefits
- 4 Universal inputs
- Accepts thermocouple, RTD, bridge based sensors, and IEPE type sensors
- Simultaneous sampling at up to 200kS/s/ch
- 16-bit resolution
- High accuracy
- Built-in counter and timer functions
- 250 VRMS or DC Cat II isolation
Compatibility
Daxus® DXS-100 and SmartCorder® DDX-100 data acquisition systems
Typical applications
- Troubleshooting
- Product testing
- Mixed-signal applications
+-Specifications
COMMON SPECIFICATIONS
- Channels Per Module: 4
- Rated Isolation: 250 VRMS or DC, CAT II (iso-common to chassis and other iso-commons)
- Frequency Counter: Capability All channels. Software selectable.
- Counter Modes: Gated time frequency counter, cycle-based frequency counter, pulse event counter, gated pulse event counter, quadrature counter, pulse width detector, period width detector, duty cycle detector, edge separation detector.
- Frequency ctr range (menu): Up to 80 KHz
- Frequency ctr range (spec’d): 2 – 40 KHz
- Frequency ctr accuracy: ± 0.07% of Measurement + .002 Hz
- Min counter input amplitude: 25% of span for freq and pulse counters, 90% of span for all other modes
- Pulse counter range: 4000000000 maximum. (16 bit display resolution)
- Pulse width accuracy: .002% of measurement + .00167% of span + 0.7 μs
- Pulse width range: 25 μs – 2500000
- Edge separation accuracy: .002% of measurement + .00167% of span + 0.7 μs
- Edge separation range: 25 μs – 5000000 μs
- Period width accuracy: .001% of measurement + .00167% of span + 0.7 μs
- Period width range :25 μs – 100000 μs (10 Hz – 40 KHz)
- Duty cycle accuracy: .5% (Inputs in the 1 Hz – 5 KHz range with 5% – 95% duty cycles)
- Counter Timebase: 50 MHZ
- Cold Start Drift: 0.1% of attenuator (60 min.)
SINGLE ENDED INPUT
- Connector: Guarded banana jacks (red/black)
- Input: Single-ended, AC/DC coupled
- Sample Rate: 200 kS/s/ch
- A/D: 16 bit SAR (one per channel)
- Anti-Aliasing Filter: 4 pole Bessel
- Bandwidth: 40 KHz (-3dB)
- AC Coupled 3dB Point: < 0.54 Hz (0.47 Hz typ)
- Off Ground Measurements: Yes
- Zero Suppression: Digital.
- Attenuator Ranges: 1, 10, 100, 200 and 400 Volt
- Measurement Ranges: ± 400 V (400 VFS or 800 VFS w/ zero offset), ± 200 V (200 VFS or 400 VFS w/ zero offset),
± 100 V (100 VFS or 200 VFS w/ zero offset), ± 10 V (10 VFS or 20 VFS w/ zero offset),
± 1 V (1 VFS or 2 VFS w/ zero offset. 0.1V min span)
- Max Rated Input: 250 Vrms or DC, Cat II
- Max Transient Input: ± 800 V peak (not to exceed 250Vrms)
- DC Accuracy (25°C): ± 0.06% of attenuator
- Overshoot: < 0.25%
- Intrinsic Noise: (pk-pk) < 0.02% of attenuator + .02% of span (400V through 10V atts)
< 0.16% of attenuator + .02% of span (1V att)
- IMR at 60 Hz: Better than -80 dB
- Min Input Impedance > 1 Megohm
DIFFERENTIAL INPUT
- Connector: 8 wire screw terminal
- Input: Differential, DC coupled
- Sample Rate: 200 kS/s/ch
- A/D: 16 bit SAR (one per channel)
- Anti-Aliasing Filter: 4 pole Bessel
- Bandwidth: 35 KHz
- Measurement Ranges: ± 1000 mV, ± 100 mV, ± 20 mV
- Max Transient Input: ± 20 V (no damage)
- Common Mode Voltage: ± 10 V
- Zero Suppression: digital.
- DC Accuracy (25°C): ± 0.06% of attenuator
- Overshoot: < 0.25%
- Intrinsic Noise: < 0.02% of attenuator + .02% of span (1000 mV Att), < 0.05% of attenuator + .02% of span (100 mV Att)
< 0.18% of attenuator + .02% of span (20 mV Att)
- Input Impedance: > 300 KΩ (150 KΩ balanced to isolated common)
- CMR at 60 Hz: > 85 dB
- Excitation: DC Voltage – adjustable, 0.1 to 10 V. 30 mA maximum
- Excitation Accuracy: 0.05 V voltage mode
RESISTANCE INPUT
(32950503 ADP-R option)
- Connector: 4 wire screw terminal
- Input: Differential, DC coupled
- Sample Rate: 2.5 samples/sec
- A/D: 24 bit SigmaDelta (one per channel)
- Anti-Aliasing Filter: Inherent
- Measurement Ranges: 0 to 1500 Ω
- Max Transient Input: ± 20 V (no damage)
- Accuracy: 0.03% of measurement + 0.04 Ω
DCVM INPUT
- Connector: 8 wire screw terminal
- Input: Differential, DC coupled
- Sample Rate: 2.5 samples/sec
- A/D: 24 bit SigmaDelta (one per channel)
- Anti-Aliasing Filter: Inherent
- Measurement Ranges: -10 mV to + 100 mV
- Max Transient Input: ± 20 V (no damage)
- Common Mode Voltage: ± 10V
- Zero Suppression: Digital.
- DC Accuracy (25°C): 20 V (no damage) 0.04% of attenuator
- Intrinsic Noise (pk-pk): < 0.005% of attenuator
- Input Impedance: > 300 KΩ (150 KΩ balanced to isolated common)
RTD INPUT
(32950503 ADP-R option)
- Connector: 4 wire screw terminal
- Input: Differential, DC coupled
- Sample Rate: 2.5 samples/sec
- A/D: 24 bit SigmaDelta (one per channel)
- Anti-Aliasing Filter: Inherent
- Measurement Ranges: Pt100(385) -200 to 800°C,
Pt100(3916) -200 to 630°C (-200 to 800 on menu)
- Max Transient Input: ± 20 V (no damage)
- Resolution: 0.01 *C
- Supported RTD Probe types: Pt 100 – 385 (DIN 43760, IEC751 and ASTM 1137), Pt 100 – 3916 (JIS C1604),
Pt 100 – 3926 (reference grade) – future
- DC Accuracy (25°C): Pt 100 – 385 0.04% of measurement + 0.1°C, Pt 100 – 3916 0.1% of measurement + 0.2°C
- Linearization: Yes
THERMOCOUPLE INPUT
(32950502 ADP-T needed for internal CJC)
- Connector: Type U miniature thermocouple
- Input: Differential, DC coupled
- Sample Rate: 2.5 samples /sec
- A/D: 24 bit SigmaDelta (one per channel)
- Anti-Aliasing Filter: Inherent
- Specified Range Type J: -210 to 1200°C
- Specified Range Type K: -200 to 1372°C
- Specified Range Type E: -200 to 1000°C
- Specified Range Type T: -200 to 400°C
- Specified Range Type N: -200 to 1300°C
- Specified Range Type B: 600 to 1820°C (250 to 1820 on menu)
- Specified Range Type R: 0 to 1767°C (-20 to 1768 on menu)
- Specified Range Type S: 0 to 1767°C (-20 to 1768 on menu)
- Specified Range Type C: 0 to 2316°C
- Max Transient Input: ± 20 V (no damage)
- Common Mode Voltage: ± 10V
- Resolution: 0.01°C
- Thermocouple types: J,K,E,T,N,B,R,S,C
- Accuracy (25°C) J (< 0): ± 3.0°C.
- Accuracy (25°C) J (0 to 1200): ± 1.0°C.
- Accuracy (25°C) K (< 0): ± 3.0°C.
- Accuracy (25°C) K (0 to 1372): ± 1.0°C.
- Accuracy (25°C) E (< -100): ± 3.0°C.
- Accuracy (25°C) E (-100 to 1000): ± 1.5°C.
- Accuracy (25°C) T (< -100): ± 2.5°C.
- Accuracy (25°C) T (-100 to 400): ± 1.5°C.
- Accuracy (25°C) N (< -50): ± 3.0°C.
- Accuracy (25°C) N (-50 to 1300): ± 1.5°C.
- Accuracy (25°C) B: ± 4.5°C.
- Accuracy (25°C) R: ± 5.5°C.
- Accuracy (25°C) S: ± 5.5°C.
- Accuracy (25°C) C: (W5ReM26Re) ± 3.0°C.
- Cold Junction Compensation: Both internal and external.
- Compensation Error: Included in accuracy specification
- Linearization: NIST ITS-90
- Input Impedance: > 300 KΩ (150 KΩ balanced to isolated common)
IEPE (LIVM) INPUT
(32950501 ADP-I option)
- Connector: Isolated BNC
- Input: Single Ended, AC coupled with constant current excitation
- Sample Rate: 200 kS/s/ch
- A/D: 16-bit SAR (one per channel)
- Anti-Aliasing Filter: 4 pole Bessel
- Measurement Ranges: ± 10 V (1 to 20 VFS)
- Max Transient Input: ± 30 V (no damage)
- Accuracy (25°C): ± 0.1% of attenuator
- Overshoot: < 0.25%
- Intrinsic Noise (pk-pk): < 0.02% of attenuator + .02% of span
- Min Input Impedance: > 1 Megohm
- IMR at 60 Hz: Better than -85 dB
- Excitation: 4.5 mA DC current
- Excitation Accuracy (25°C): 20%
- Excitation Compliance Voltage: 20V (Approx 24V open circuit voltage)
- Excitation Protection: Short Circuit Protected
- Coupling Time Constant: 10 s (± 30%) (10μF / 1M‡)
- Rated Isolation: 30 Vrms or 60 VDC
Optional accessories
| SC-DAX Soft Carry Case for Daxus DXS-100, P/N: 42737000 |
| HC-DXS Hard Carry Case for Daxus DXS-100, P/N: 41047300 |
| SC-DDX Soft Carry Case for SmartCorder DDX-100, P/N: 41047200 |
Available services
- 1-Year extended warranty
- Factory calibration service